DNA Isolation up to 2 mL of Whole Blood

Whole blood samples are critical for disease prediction and diagnosis. Obtaining intact, high quality and concentrated genomic DNA (gDNA) from whole blood is the critical step for downstream research applications such as qPCR, microarray and next generation sequencing (NGS). The term biobank refers to a large collection of tissue samples such as blood and serum that are collected for research purposes. To have access to large amounts of human samples and data, an increasing number of researchers are looking to access available samples through biobanks. Here we present a high-quality, research-ready DNA extraction method from large volume whole blood by using GenFind V3.

Protocol:

DNA extraction up to 2 mL blood using GenFind V3

DNA Extraction up to 2 mLs of Blood Collected in EDTA, Heparin and Citrate Blood Collection Tubes using GenFind V3

Isolation of genomic DNA from 2 mL frozen whole blood was demonstrated by isolating DNA from four healthy blood donors. The blood was collected in three blood collection tubes (BCT) that contained the following preservatives: EDTA, heparin and citrate. DNA yield and purity was accessed using a NanoDrop (Thermo Fisher Scientific). The average DNA yield was 33 μg for EDTA collection tubes, 25 μg for heparin collection tubes and 23 μg for citrate collection tubes (Figure 1).

Genomics PoP DNA Isolation from 2mL Whole Blood Figure 1

 

The average A260/A280 purity ratio was 1.7 for all collection tubes (Figure 2).

Genomics PoP DNA Isolation from 2mL Whole Blood Figure 2

 

The average A260A/230 purity ratio was 1.9 for all collection tubes (Figure 3).

Genomics PoP DNA Isolation from 2mL Whole Blood Figure 3

 

Beckman Coulter makes no warranties of any kind whatsoever express or implied, with respect to this protocol, including but not limited to warranties of fitness for a particular purpose or merchantability or that the protocol is non-infringing. All warranties are expressly disclaimed. Your use of the method is solely at your own risk, without recourse to Beckman Coulter. Not intended or validated for use in the diagnosis of disease or other conditions. This protocol is for demonstration only, and is not validated by Beckman Coulter.
© 2022 Beckman Coulter, Inc. All rights reserved. Beckman Coulter, the stylized logo and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other counters. All other trademarks are the property of their respective owners.